Model CI-XT60NProposal 202610P01Quotation 261015Q02
Chang Ai Technology · One Gasmaster, Petaling Jaya
Online Flue Gas Analysis System for Biomass Boiler
Floor-standing analysis cabinet, 600 W × 600 D × 2000 H mm on a 100 mm plinth, for continuous O₂, CO and CO₂ monitoring of biomass boiler flue gas on a dry basis. Three 19-inch analyzers sit behind the door window. Below them, the sampling panel carries the precision filter, sample gas pressure gauge and regulating valve, calibration port, one flowmeter per analyzer and the sample / calibration valve, with the CI-R300P sample gas cooler underneath.
Drag to rotate · scroll or pinch to zoom · right-drag to pan
Door
Rear door
View
Measurement requirements
Parameter
Analyzer
Principle
Measuring range
Oxygen O₂
CI-PC912
Electrochemical
0~25.00 vol%
Carbon Monoxide CO
CI-PC21-C
Enhanced NDIR
0~10,000 ppm
Carbon Dioxide CO₂
CI-PC21-B
Enhanced NDIR
0~25.00 vol%
Process: biomass boiler flue gas at about 350 °C. Results reported on a dry basis after the compressor condenser.
Analyzer performance
Analyzer
Accuracy
Repeatability
Drift
Response
Sample pressure
Sample flow
CI-PC912 O₂
±2 % F.S.
±1 % F.S.
±1 % F.S. / 7 d
T90 < 20 s
Atmospheric ±10 %
500~1000 mL/min
CI-PC21-C CO
±2 % F.S.
±1 % F.S.
Zero and span ±1 % F.S. / 7 d
T90 ≤ 60 s at 0.5 L/min
Atmospheric ±10 %
500~1000 mL/min
CI-PC21-B CO₂
±2 % F.S.
±1 % F.S.
Zero and span ±1 % F.S. / 7 d
T90 ≤ 60 s at 0.5 L/min
Atmospheric ±10 %
500~1000 mL/min
CI-PC21 inlet limits: particles ≤ 1 µm and ≤ 200 µg/Nm³, water content at saturation no higher than 5 °C with no condensation, which sets the duty of the condenser and filter. CI-PC21 case is 19-inch, 483 × 300 × 177 mm; CI-PC912 is shown in the same 19-inch case style.
Cabinet
Model
CI-XT60N
Dimensions
600 W × 600 D × 2000 H mm, plus 100 mm plinth
Enclosure
Carbon steel, blasted and powder-coated, RAL 7035; canopy roof with four lifting eyes
Doors
Front door with round-cornered window and black rubber lining; rear door with ventilation louvres. Both right-hinged, lockable handles aligned
Installation
Indoor or sheltered, ventilated location; no cabinet air conditioner
Panel layout
Name plate with power indicator; three 19-inch analyzers, each with a label strip; sampling panel; sample gas cooler; lower cover over the electrical compartment
Connections
Heated sample line, purge air, vent, drain, power and signal entries at the cabinet floor
Signals and power
Analog output
One isolated 4~20 mA per analyzer (3 channels), selectable 0~20 mA / 0~10 V / 1~5 V
Communication
RS485 Modbus RTU (standard), RS232 (optional)
Relays
2 concentration alarms per analyzer, 24 VDC 0.2 A contacts
Control input
9~28 VDC external control signals
Power
AC 220 V, 50 Hz to the cabinet; analyzers accept 85~264 VAC
Sample path
High-temperature sampling probe with filter and automatic back-purge, at the flue duct
Heated sample line, 20 m, held above the flue gas dew point
CI-R300P sample gas cooler (compressor condenser); condensate removed by the peristaltic pump
GF2800 precision filter
Sample pump, then sample gas pressure gauge and pressure regulating valve
3-way valve, sample / calibration, fed by the front calibration port
One flowmeter per analyzer, then CI-PC912, CI-PC21-C and CI-PC21-B
Analyzer outlets to the common vent at the cabinet floor
Calibration and maintenance
Cycle
1 year recommended for each analyzer
Zero gas
High-purity nitrogen
Span gas
CO and CO₂ in N₂, and O₂ in N₂, at about 80 % of each range
Gas supply
Calibration gases supplied by the buyer
Routine
Probe and GF2800 filter elements, peristaltic pump tubing, condensate drain check, leak check of fittings
Concept model for proposal illustration. Final internal layout follows the factory general arrangement drawing.