Model CI-XT60N-MOProposal 202609P01Quotation 260901Q02
Chang Ai Technology · Mora Pharmaceuticals & Consulting, Buenos Aires
Online Medical Gas Analysis System
Floor-standing 19-inch analysis cabinet for continuous quality monitoring of medicinal oxygen to Ph. Eur. and USP limits. Four rack analyzers measure O₂, CO, CO₂ and H₂O; the panel below handles pressure reduction, flow control and calibration.
Drag to rotate · scroll or pinch to zoom · right-drag to pan
Door
Rear door
View
Measurement requirements
Parameter
Analyzer
Ph. Eur.
USP
Measuring range
Oxygen O₂
CI-PC312
≥ 99.5 % v/v
≥ 99.0 % v/v
0~99.99 %
Carbon Monoxide CO
CI-PC21-C
< 5 ppm v/v
< 10 ppm v/v
0~20 ppm
Carbon Dioxide CO₂
CI-PC21-B
< 300 ppm v/v
< 300 ppm v/v
0~1000 ppm
Moisture H₂O
CI-PC31
< 67 ppm v/v
Not specified
−100~+20 °C Td
Analyzer performance
Analyzer
Principle
Accuracy
Repeatability
Resolution
LOD
Response
Sample flow
CI-PC312 O₂
Paramagnetic
±0.1 % (linearity)
—
0.01 %
0.05 %
T10~90 < 2.5 s
500~1000 mL/min
CI-PC21-C CO
NDIR
±2 % F.S.
±1 % F.S.
0.01 ppm
0.1 ppm
T90 < 15 s
500~1000 mL/min
CI-PC21-B CO₂
NDIR
±2 % F.S.
±1 % F.S.
0.1 ppm
0.5 ppm
T90 < 15 s
500~1000 mL/min
CI-PC31 H₂O
Aluminium oxide capacitive
±2–3 °C
±0.5 °C
0.1 °C
−100 °C
5~45 s
2~3 L/min
Sample pressure at each analyzer: atmospheric ±10 %. Expected sensor life > 2 years, analyzer life > 5 years under normal use.
Cabinet
Model
CI-XT60N-MO
Dimensions
600 W × 1800 H × 800 D mm, on 100 mm base
Enclosure
19-inch standard analysis cabinet, powder-coated steel, RAL 7035
Doors
Front door with round-cornered window and thick black rubber lining; solid rear door. Both right-hinged (seen from the front) with lockable handles aligned
Rack layout
CI-PC312 3U; CI-PC21-C, CI-PC21-B, CI-PC31 4U each; 1U label strip under each analyzer
Connections
Sample inlet and cables at bottom, exhaust outlet at top, calibration gas inlets on front panel
Lifting
4 × eye bolts on roof
Signals and data
Communication
RS485 Modbus (standard), RS232 (optional)
Analog output
4~20 mA, selectable 0~20 mA / 0~1 V / 0~5 V / 0~10 V / 1~5 V
Sample inlet ON/OFF valve, pressure regulator and gauge: 0.6–20 MPa reduced to atmospheric
Dust filter and gas cooling, mounted behind the panel
Flowmeters with needle valves: bypass, O₂, CO, CO₂, H₂O
Calibration lines for CI-PC312, CI-PC21-C and CI-PC21-B, each with ON/OFF valve, regulator and gauge
3-way selector valves (sample / calibration) with front-panel calibration inlets
CI-PC31 sample ON/OFF valve
Exhaust header to the top outlet; SS316 tubing and fittings routed behind the panel
Calibration and maintenance
Cycle
1 year recommended, all analyzers
Zero gas
5N nitrogen, 99.9995 % N₂
Span gas
O₂ 80.0 % · CO 16 ppm · CO₂ 800 ppm · H₂O not applicable
Gas supply
Calibration gases sourced locally in Argentina by the buyer
Certificates
Factory calibration certificate with each analyzer
Routine
Filter element replacement, leak check of gas-path fittings
Concept model for proposal illustration, based on the CI-XT60N1292 cabinet and panel arrangement. Final internal layout follows the factory general arrangement drawing.